← RecallGuard
HomeRecalls

Beckman Coulter, Inc. Recall: DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Numb

Beckman Coulter, Inc. Recall: DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Numb is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Beckman Coulter, Inc.; Product: DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457; Reason for recall: Beckman Coulter has ident

Recalling firmBeckman Coulter, Inc.
ProductDxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457
Reason for recallBeckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionWorldwide distribution - US Nationwide.
Recall initiated20250625
Recall numberZ-2267-2025
Categorydevice
StateMN
CountryUnited States

🔍 Search all recalls →