Beckman Coulter, Inc. Recall: DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Numb
Beckman Coulter, Inc. Recall: DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Numb is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Beckman Coulter, Inc.; Product: DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457; Reason for recall: Beckman Coulter has ident
| Recalling firm | Beckman Coulter, Inc. |
|---|---|
| Product | DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457 |
| Reason for recall | Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | Worldwide distribution - US Nationwide. |
| Recall initiated | 20250625 |
| Recall number | Z-2267-2025 |
| Category | device |
| State | MN |
| Country | United States |