Beckman Coulter Inc. Recall: DxM 1096 MicroScan WalkAway Instrument REF B1018-496
Beckman Coulter Inc. Recall: DxM 1096 MicroScan WalkAway Instrument REF B1018-496 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Beckman Coulter Inc.; Product: DxM 1096 MicroScan WalkAway Instrument REF B1018-496; Reason for recall: WalkAway Instruments may have been assembled with th
| Recalling firm | Beckman Coulter Inc. |
|---|---|
| Product | DxM 1096 MicroScan WalkAway Instrument REF B1018-496 |
| Reason for recall | WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | Worldwide distribution - U.S. Nationwide distribution in the states of Alaska, Arkansas, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Missouri, New Jersey, New Mexico, New York, North Carolina |
| Recall initiated | 20191021 |
| Recall number | Z-1086-2022 |
| Category | device |
| State | CA |
| Country | United States |