Beckman Coulter Inc. Recall: MicroScan WalkAway-40 plus Instrument REF B1018-283

Beckman Coulter Inc. Recall: MicroScan WalkAway-40 plus Instrument REF B1018-283 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Beckman Coulter Inc.; Product: MicroScan WalkAway-40 plus Instrument REF B1018-283; Reason for recall: WalkAway Instruments may have been assembled with the

Recalling firmBeckman Coulter Inc.
ProductMicroScan WalkAway-40 plus Instrument REF B1018-283
Reason for recallWalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionWorldwide distribution - U.S. Nationwide distribution in the states of Alaska, Arkansas, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Missouri, New Jersey, New Mexico, New York, North Carolina
Recall initiated20191021
Recall numberZ-1083-2022
Categorydevice
StateCA
CountryUnited States