Beckman Coulter Inc. Recall: UIBC (Unsaturated Iron Binding Capacity), REF: OSR61205,

Beckman Coulter Inc. Recall: UIBC (Unsaturated Iron Binding Capacity), REF: OSR61205, is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Beckman Coulter Inc.; Product: UIBC (Unsaturated Iron Binding Capacity), REF: OSR61205,; Reason for recall: Beckman Coulter identified that the Unsatura

Recalling firmBeckman Coulter Inc.
ProductUIBC (Unsaturated Iron Binding Capacity), REF: OSR61205,
Reason for recallBeckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL hemoglobin as indicated in their instructions for use. Initial internal testing confirmed that samples with low UIBC levels failed the claimed hemolysis interference specifications with reported biases up to -43.6 % UIBC when hemolyzed serum samples contained 200 mg/dL hemoglobin.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionUS: AK, AL, AR, CO, CT, DC, DE, ID, UT OUS: Algeria, Argentina, Australia, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Colombia, Costa Rica, Croatia, Czechia, Egypt, El Salvador, Ethiopia, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hung
Recall initiated20250626
Recall numberZ-2264-2025