← RecallGuard
HomeRecalls

Becton Dickinson & Co. Recall: EpiCenter Plus Microbiology Data Management System software

Becton Dickinson & Co. Recall: EpiCenter Plus Microbiology Data Management System software is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Becton Dickinson & Co.; Product: EpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the MALDI

Recalling firmBecton Dickinson & Co.
ProductEpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the MALDI BioTyper System . Product Usage: The EpiCenter system is a clinical data management system designed for use with certain BD instrument systems. The EpiCenter system provides advanced capabilities in data management, including extended demographic data, comprehensive reporting features including a powerful database query tool, bi-directional Laboratory Information System (LIS) communications and long-term data storage. Additionally, the EpiCenter system provides an Expert System th
Reason for recallThe information in the EpiCenter interface translation tables was not updated to match the Bruker MALDI BioTyper database identifications for some organisms. This could lead to a possible mis-identification of an organism.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
Distribution15 US customers, 18 Ex-US sites, and 0 distributors. Approximately 115 customers totally and 0 distributors
Recall initiated20140825
Recall numberZ-1414-2017
Categorydevice
StateMD
CountryUnited States

🔍 Search all recalls →