Belmont Instrument LLC Recall: 3.0 Liter Disposable Set (4 sets/case). Model Number: 903-0

Belmont Instrument LLC Recall: 3.0 Liter Disposable Set (4 sets/case). Model Number: 903-0 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Belmont Instrument LLC; Product: 3.0 Liter Disposable Set (4 sets/case). Model Number: 903-00032. The P/N 903-00032 consists of 4 Large Volume Rese

Recalling firmBelmont Instrument LLC
Product3.0 Liter Disposable Set (4 sets/case). Model Number: 903-00032. The P/N 903-00032 consists of 4 Large Volume Reservoirs (P/N 903-00018) and 4 Heat Exchanger/Patient Line set (P/N 403-00382). The two components are packaged separately in individual pouches and placed in the same case for shipment. The female quick connector that may be cracked is located on the Heat Exchanger/Patient Line set.
Reason for recallPotential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack during priming of the Rapid Infuser.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionNationwide distribution to AL, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NJ, NV, NY, OH, OR, PA, SC, TX, UT, VA, WA, WI. International distribution to Czechia, Italy, Norway.
Recall initiated20250331
Recall numberZ-1678-2025
Categorydevice
StateMA
CountryUnited States