Bien-Air Surgery Sa Recall: Bien-Air Surgery REF 1600517-001 SN ******* MULTIFUNCTION PEDA
Bien-Air Surgery Sa Recall: Bien-Air Surgery REF 1600517-001 SN ******* MULTIFUNCTION PEDA is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Bien-Air Surgery Sa; Product: Bien-Air Surgery REF 1600517-001 SN ******* MULTIFUNCTION PEDAL OSSEODUO - Product Usage: Product Usage: intended to
| Recalling firm | Bien-Air Surgery Sa |
|---|---|
| Product | Bien-Air Surgery REF 1600517-001 SN ******* MULTIFUNCTION PEDAL OSSEODUO - Product Usage: Product Usage: intended to convert electrical energy to mechanically drive attachment, microsaws, shaver blades and burs, burs and perforator tools for cutting and shaping bone including spine and cranium. |
| Reason for recall | The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | US Nationwide distribution including in the states of CA, MI and MN. |
| Recall initiated | 20200911 |
| Recall number | Z-0177-2021 |
| Category | device |
| State | N/A |
| Country | Switzerland |