← RecallGuard
HomeRecalls

Bio-Rad Laboratories, Inc. Recall: D-10" Hemoglobin Testing System, Model number: 220-0220

Bio-Rad Laboratories, Inc. Recall: D-10" Hemoglobin Testing System, Model number: 220-0220 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Bio-Rad Laboratories, Inc.; Product: D-10" Hemoglobin Testing System, Model number: 220-0220 Manufactured by: Bio-Rad Laboratories, Inc. 4000 Alfre

Recalling firmBio-Rad Laboratories, Inc.
ProductD-10" Hemoglobin Testing System, Model number: 220-0220 Manufactured by: Bio-Rad Laboratories, Inc. 4000 Alfred Nobel Drive Hercules, CA 94545 The D-10 Hemoglobin Testing System is a single module designed for hemoglobin analysis in mid-volume laboratories. The D-10 is a fully integrated system for sample dilution, processing and analysis of hemoglobin, designed for use with specific Bio-Rad reagent kits. The D-10 incorporates the use of a dedicated software system for instrument control, data collection and analysis.
Reason for recallOn a rare occasion the D-10 software might generate an extra result by duplicating the result form the previous sample and the Sample ID/Injection number from the next sample.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide Distribution - USA (nationwide) and Internationally to Canada, Japan, China, India, SriLanka, Argentina, Paraguay, Uruguay, Chile, Peru ,Bolivia ,Mexico ,Panama, ,Costa Rica ,Ecuador, Venezuela, Colombia, Brazil, Jamaica, Austria, Switzerland, France, Germany, Greece, Portugal, Spain, Ital
Recall initiated20140227
Recall numberZ-1299-2014
Categorydevice
StateCA
CountryUnited States

🔍 Search all recalls →