Biomet, Inc. Recall: REF 154225Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Si
Biomet, Inc. Recall: REF 154225Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Si is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Biomet, Inc.; Product: REF 154225Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size D7 ARCOM UHMWPE / CO-CR-MO Alloy Only for use in Lat
| Recalling firm | Biomet, Inc. |
|---|---|
| Product | REF 154225Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size D7 ARCOM UHMWPE / CO-CR-MO Alloy Only for use in Lateral compartment Sterile R The Oxford Fixed Bearing Knee is a Partial knee replacement system consisting of a femoral component and a fixed bearing tibial component for use in the lateral compartment of the knee. The Oxford Fixed Lateral bearings are indicated for use in the lateral compartment and intended to be implemented with bone cement. |
| Reason for recall | Biomet investigation found the design of the tibial bearing creates the potential for anterior impingement between the bone and the poly bearing component in terminal extension. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Worldwide distribution: USA including NC, IN, PA, and OH; and countries of: Canada and Netherlands. |
| Recall initiated | 20130607 |
| Recall number | Z-1714-2013 |
| Category | device |
| State | IN |
| Country | United States |
$29/mo
Try RecallGuard →