Biosense Webster, Inc. Recall: 20-Pole LASSO Nav Catheter Eco, Catalog # D-1349-02S. Desig
Biosense Webster, Inc. Recall: 20-Pole LASSO Nav Catheter Eco, Catalog # D-1349-02S. Desig is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Biosense Webster, Inc.; Product: 20-Pole LASSO Nav Catheter Eco, Catalog # D-1349-02S. Designed to facilitate electrophysiological mapping of the a
| Recalling firm | Biosense Webster, Inc. |
|---|---|
| Product | 20-Pole LASSO Nav Catheter Eco, Catalog # D-1349-02S. Designed to facilitate electrophysiological mapping of the atria of the heart. |
| Reason for recall | Biosense Webster is recalling the 20-Pole LASSO NAV Catheter Eco because it has been incorrectly calibrated causing them to be recognized and displayed as a 10-Pole LASSO NAV Catheter Eco by the CARTO 3 System. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Distributed in the states of MN, FL, and UT. |
| Recall initiated | 20130619 |
| Recall number | Z-1617-2013 |
| Category | device |
| State | CA |
| Country | United States |