← RecallGuard
HomeRecalls

Biosense Webster, Inc. Recall: 20-Pole LASSO Nav Catheter Eco, Catalog # D-1349-02S. Desig

Biosense Webster, Inc. Recall: 20-Pole LASSO Nav Catheter Eco, Catalog # D-1349-02S. Desig is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Biosense Webster, Inc.; Product: 20-Pole LASSO Nav Catheter Eco, Catalog # D-1349-02S. Designed to facilitate electrophysiological mapping of the a

Recalling firmBiosense Webster, Inc.
Product20-Pole LASSO Nav Catheter Eco, Catalog # D-1349-02S. Designed to facilitate electrophysiological mapping of the atria of the heart.
Reason for recallBiosense Webster is recalling the 20-Pole LASSO NAV Catheter Eco because it has been incorrectly calibrated causing them to be recognized and displayed as a 10-Pole LASSO NAV Catheter Eco by the CARTO 3 System.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionDistributed in the states of MN, FL, and UT.
Recall initiated20130619
Recall numberZ-1617-2013
Categorydevice
StateCA
CountryUnited States

🔍 Search all recalls →