← RecallGuard
HomeRecalls

Capintec Inc Recall: Captus 4000e Thyroid Uptake System, Model Numbers 5430-30151 and 5430

Capintec Inc Recall: Captus 4000e Thyroid Uptake System, Model Numbers 5430-30151 and 5430 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Capintec Inc; Product: Captus 4000e Thyroid Uptake System, Model Numbers 5430-30151 and 5430-30152 - Product Usage: is intended to be used by train

Recalling firmCapintec Inc
ProductCaptus 4000e Thyroid Uptake System, Model Numbers 5430-30151 and 5430-30152 - Product Usage: is intended to be used by trained nuclear Medicine Technologists or Nuclear Medicine Physicians to perform Thyroid Uptake Procedures.
Reason for recallIt is possible that the collimators have screw hole locations that were insufficiently tapped. There is a potential for the collimator assembly to fall and come into contact with a patient or operator.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide distribution. US Nationwide, Australia, Austria, Bangladesh, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, France, Hong Kong, India, Israel, Italy, Korea, Malaysia, Pakistan, Philippines, Poland, Saudi Arabia, Spain, Sweden, Taiwan, Thailand, Turkey, UAE, UK, and Vietnam. (list upd
Recall initiated20200303
Recall numberZ-1564-2020
Categorydevice
StateNJ
CountryUnited States

🔍 Search all recalls →