← RecallGuard
HomeRecalls

CareFusion 303, Inc. Recall: Alaris PC unit model 8015 Product Usage: The device is labele

CareFusion 303, Inc. Recall: Alaris PC unit model 8015 Product Usage: The device is labele is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: CareFusion 303, Inc.; Product: Alaris PC unit model 8015 Product Usage: The device is labeled for prescription use only.; Reason for recall: The re

Recalling firmCareFusion 303, Inc.
ProductAlaris PC unit model 8015 Product Usage: The device is labeled for prescription use only.
Reason for recallThe recall was initiated because the Alaris PC unit model 8015 has a component on the PC unit power supply board that is causing an error code(120.4630) "System Error" or "Missing Battery Error" at start up. The error code is accompanied by both an audible alarm and visual error messages on the PC unit screen.
ClassificationClass I (most serious — reasonable probability of serious health consequences or death)
StatusTerminated
DistributionUS Nationwide Distribution
Recall initiated20120525
Recall numberZ-2076-2012
Categorydevice
StateCA
CountryUnited States