← RecallGuard
HomeRecalls

CareFusion 303, Inc. Recall: BD Pyxis MedBank MedPass Software, REF: 139088-01

CareFusion 303, Inc. Recall: BD Pyxis MedBank MedPass Software, REF: 139088-01 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: CareFusion 303, Inc.; Product: BD Pyxis MedBank MedPass Software, REF: 139088-01; Reason for recall: After dispensing a timed dose medication for a patient, th

Recalling firmCareFusion 303, Inc.
ProductBD Pyxis MedBank MedPass Software, REF: 139088-01
Reason for recallAfter dispensing a timed dose medication for a patient, the automated dispensing cabinet software may continue to show the already dispensed medication dose is still available for dispense, which could lead to duplicate medication administration.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionUS: FL, MI, KS
Recall initiated20230818
Recall numberZ-2646-2023
Categorydevice
StateCA
CountryUnited States

🔍 Search all recalls →