Codman & Shurtleff, Inc. Recall: Codman Lumbar Drainage Catheter Kit II with EDS 3 (Extern

Codman & Shurtleff, Inc. Recall: Codman Lumbar Drainage Catheter Kit II with EDS 3 (Extern is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Codman & Shurtleff, Inc.; Product: Codman Lumbar Drainage Catheter Kit II with EDS 3 (External Drainage System) Catalog Number: 82-1738 Indicated f

Recalling firmCodman & Shurtleff, Inc.
ProductCodman Lumbar Drainage Catheter Kit II with EDS 3 (External Drainage System) Catalog Number: 82-1738 Indicated for draining cerebrospinal fluid (CFS) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume when the insertion of a permanent, internal shunt is not indicated
Reason for recalltubing within the system that drains CSF may leak or disconnect from the joints and may result in over- or under-drainage of CSF from the ventricular system or introduction of air into the ventricular system (pneumocephalus)
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide Distribution - USA (nationwide) and Internationally to: United Arab Emirates Argentina Austria Australia Aruba Belgium Bahrain Bolivia Brazil Bahamas Canada Switzerland Chile People's Republic of China Columbia Costa Rica Cyprus Czech Republic Germany Algeria Estonia Egypt Spain Finland Fr
Recall initiated20140915
Recall numberZ-0091-2015
Categorydevice
StateMA
CountryUnited States