Cook Inc. Recall: Flexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL0-HC, Order Nu
Cook Inc. Recall: Flexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL0-HC, Order Nu is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Cook Inc.; Product: Flexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL0-HC, Order Number G56231 Flexor Introducers and Guiding Sheaths are inte
| Recalling firm | Cook Inc. |
|---|---|
| Product | Flexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL0-HC, Order Number G56231 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature. |
| Reason for recall | There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage. |
| Classification | Class I (most serious — reasonable probability of serious health consequences or death) |
| Status | Terminated |
| Distribution | US Nationwide Distribution including states of: AK, CT, IA, LA, MO, NH, OK, TN, WI, AL, DC, ID, MA, MS, NJ, OR, TX, WV, AR, DE, IL, MD, MT, NM, PA, UT, WY, AZ, FL, IN, ME, NC, NV, RI, VA, CA, GA, KS, MI, ND, NY, SC, VT, CO, HI, KY, MN, NE, OH, SD, WA and Global Distribution to: AU, SG, Austria, Belg |
| Recall initiated | 20201124 |
| Recall number | Z-0598-2021 |
| Category | device |
| State | IN |
| Country | United States |