Datascope Corporation Recall: BEQ-TOP 22702 ECC W/HMOD 30000, Catalog No. 701051179
Datascope Corporation Recall: BEQ-TOP 22702 ECC W/HMOD 30000, Catalog No. 701051179 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Datascope Corporation; Product: BEQ-TOP 22702 ECC W/HMOD 30000, Catalog No. 701051179; Reason for recall: The Better Bladder device provided as part of so
| Recalling firm | Datascope Corporation |
|---|---|
| Product | BEQ-TOP 22702 ECC W/HMOD 30000, Catalog No. 701051179 |
| Reason for recall | The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Distributed to accounts in AL, AR, CA, DC, FL, IL, LA, MO, NC, PA, SC, TX, and WA. |
| Recall initiated | 20171019 |
| Recall number | Z-1712-2018 |
| Category | device |
| State | NJ |
| Country | United States |