← RecallGuard
HomeRecalls

Degania Silicone, Ltd. Recall: Temperature sensor Foley catheter 10FR, Catalogue Number 10

Degania Silicone, Ltd. Recall: Temperature sensor Foley catheter 10FR, Catalogue Number 10 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Degania Silicone, Ltd.; Product: Temperature sensor Foley catheter 10FR, Catalogue Number 102205101080AB Product Usage: This is sterile, single use

Recalling firmDegania Silicone, Ltd.
ProductTemperature sensor Foley catheter 10FR, Catalogue Number 102205101080AB Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
Reason for recallThe catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
Recall initiated20180621
Recall numberZ-2756-2018
Categorydevice
StateN/A
CountryIsrael

🔍 Search all recalls →