DePuy Orthopaedics, Inc. Recall: SOL SYS 8IN CALC 1.5/18.0 Packaging: Product is packed wi

DePuy Orthopaedics, Inc. Recall: SOL SYS 8IN CALC 1.5/18.0 Packaging: Product is packed wi is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: DePuy Orthopaedics, Inc.; Product: SOL SYS 8IN CALC 1.5/18.0 Packaging: Product is packed within a poly protector, with protection added as needed

Recalling firmDePuy Orthopaedics, Inc.
ProductSOL SYS 8IN CALC 1.5/18.0 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam sleeve for added protection. The assembly is then placed and sealed in an inner and outer tray for aseptic transfer. The sealed package is placed into a folding carton, labeled, and shrink wrapped. Intended Use: SOLUTION System Calcar Stems are used in total hip arthroplasty (THA) to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of
Reason for recallAn Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionNationwide Distribution including AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY and GUAYNABO, PR
Recall initiated20121203
Recall numberZ-0577-2013
Categorydevice
StateIN
CountryUnited States