Draeger Medical, Inc. Recall: Evita Infinity V500 Ventilator; Babylog Infinity VN500 Venti is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Draeger Medical, Inc.; Product: Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.; Reason for recall: The battery cap
| Recalling firm | Draeger Medical, Inc. |
|---|---|
| Product | Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation. |
| Reason for recall | The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device. |
| Classification | Class I (most serious — reasonable probability of serious health consequences or death) |
| Status | Terminated |
| Distribution | US Nationwide Distribution including Puerto Rico. |
| Recall initiated | 20160204 |
| Recall number | Z-0827-2016 |
| Category | device |
| State | PA |
| Country | United States |