← RecallGuard
HomeRecalls

Elekta, Inc. Recall: Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment

Elekta, Inc. Recall: Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Elekta, Inc.; Product: Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11,3, 11.4 Product D

Recalling firmElekta, Inc.
ProductBrand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11,3, 11.4 Product Description: Leksell GammaPlan¿ is a computer-based system designed for Leksell Gamma Knife¿ treatment planning.
Reason for recallDue to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionWorldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, ST, FL, GA, HI, IA, IL, LA, MA, MI, MN. MO, NC, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Algeria, Argentina, Australia, Belarus, Belgium, Brazil, Bulgaria, Canada, Chin
Recall initiated20251212
Recall numberZ-1340-2026
Categorydevice
StateGA
CountryUnited States

🔍 Search all recalls →