Elekta, Inc. Recall: Medical Linear Accelerator

Elekta, Inc. Recall: Medical Linear Accelerator is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Elekta, Inc.; Product: Medical Linear Accelerator; Reason for recall: There have been instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become

Recalling firmElekta, Inc.
ProductMedical Linear Accelerator
Reason for recallThere have been instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionAL, AZ, CA, FL, GA, KY, MI, MN, MO, NE, NJ, NY, NC, OH, OK, PA, SC, TX, WA, WV, Argentina, Austria, Australia, Bosnia and Herzegovinia, Bangladesh, Belgium, Bulgaria, Brazil, China, Columbia, Cuba, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, Indonesia, Ireland, Israe
Recall initiated20160708
Recall numberZ-2721-2016
Categorydevice
StateGA
CountryUnited States

🔍 Search all recalls →