Elekta, Inc. Recall: Precise Digital Accelerator Intended to be used for radiation therapy is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Elekta, Inc.; Product: Precise Digital Accelerator Intended to be used for radiation therapy treatments of malignant neoplastic diseases, as determ
| Recalling firm | Elekta, Inc. |
|---|---|
| Product | Precise Digital Accelerator Intended to be used for radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner. |
| Reason for recall | Elekta has identified a batch of securing bolts that fix the Gantry counter frame weight stack may fail. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Worldwide Distribution - USA including AZ, CA, FL, KS, LA, MN, MO, NY, NC and Internationally to Austria, Bulgaria, Canada, China, Germany, Hungary, Italy, Malaysia, Malta, Netherlands, Norway, Russia, Spain, Sweden, Taiwan and Ukraine. |
| Recall initiated | 20150107 |
| Recall number | Z-1088-2015 |
| Category | device |
| State | GA |
| Country | United States |