Ethicon Endo-Surgery Inc Recall: Straight Intraluminal Staplers, 33 mm diameter, Model SDH

Ethicon Endo-Surgery Inc Recall: Straight Intraluminal Staplers, 33 mm diameter, Model SDH is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Ethicon Endo-Surgery Inc; Product: Straight Intraluminal Staplers, 33 mm diameter, Model SDH33A The Ethicon Endo-Surgery Intraluminal Staplers (ILS

Recalling firmEthicon Endo-Surgery Inc
ProductStraight Intraluminal Staplers, 33 mm diameter, Model SDH33A The Ethicon Endo-Surgery Intraluminal Staplers (ILS) are anastomotic staplers available in four sizes to permit proper matching of instrument to diameter of the lumen. The instrument permits tissue attachment to the anvil shaft in a location remote from the main body of the instrument to improve access and visibility.
Reason for recallThe staplers may have an insufficient firing stroke to break the washer and completely form staples.
ClassificationClass I (most serious — reasonable probability of serious health consequences or death)
StatusCompleted
DistributionWorldwide Distribution: US (nationwide) and countries of: Argentina, Aruba, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Ecuador, Egypt, El Salvador, Guatemala, Honduras, India, Indonesia, Israel, Jamaica, Japan, Jordan, Korea, Kuwait, Lebanon, Malaysia,
Recall initiated20190411
Recall numberZ-1278-2019
Categorydevice
StateOH
CountryUnited States