Event Medical LTD Recall: Inspiration Ventilator System, i Series and LS 12.1 LCD. The Ins
Event Medical LTD Recall: Inspiration Ventilator System, i Series and LS 12.1 LCD. The Ins is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Event Medical LTD; Product: Inspiration Ventilator System, i Series and LS 12.1 LCD. The Inspiration 7i / 5i Ventilator System is to be used by hea
| Recalling firm | Event Medical LTD |
|---|---|
| Product | Inspiration Ventilator System, i Series and LS 12.1 LCD. The Inspiration 7i / 5i Ventilator System is to be used by healthcare professionals in hospitals or healthcare facilities and intra-hospital transport. |
| Reason for recall | A component on the systems power board can fail in a manner which can cause the ventilator to shut down without alarming. |
| Classification | Class I (most serious — reasonable probability of serious health consequences or death) |
| Status | Terminated |
| Distribution | Worldwide Distribution - US (Nationwide) and Internationally to: Algeria Argentina Australia Austria Bahrain Bangladesh Barbados Belarus Belgium Bolivia Canada Chile China Colombia Czech Rep Ecuador Egypt England Estonia Finland Germany Ghana Gibraltar Greece Guatemala Holland Hong Kong Hungary Indi |
| Recall initiated | 20151013 |
| Recall number | Z-0230-2016 |
| Category | device |
| State | CA |
| Country | United States |