← RecallGuard
HomeRecalls

Exactech, Inc. Recall: Exactech Equinoxe GLENOID, POSTERIOR AUGMENT,PEGGED, CEMENTED, 8o,

Exactech, Inc. Recall: Exactech Equinoxe GLENOID, POSTERIOR AUGMENT,PEGGED, CEMENTED, 8o, is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Exactech, Inc.; Product: Exactech Equinoxe GLENOID, POSTERIOR AUGMENT,PEGGED, CEMENTED, 8o, RIGHT, Mates with: a) 38, 41, 44, 47 head, Small, Item N

Recalling firmExactech, Inc.
ProductExactech Equinoxe GLENOID, POSTERIOR AUGMENT,PEGGED, CEMENTED, 8o, RIGHT, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-02-32, b) 41, 44, 47, 50 head, Medium, Item Number 314-02-33, c) 44, 47, 50, 53 head, Large, Item Number 314-02-34, d) 44, 47, 50, 53 head, Extra Large, Item Number 314-02-35; Shoulder Arthroplasty
Reason for recallThe packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionNationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand
Recall initiated20240306
Recall numberZ-1407-2024
Categorydevice
StateFL
CountryUnited States

🔍 Search all recalls →