← RecallGuard
HomeRecalls

Exactech, Inc. Recall: EXACTECH Equinoxe Narrow Modular Reverse Pilot-Tip Starter Reamer C

Exactech, Inc. Recall: EXACTECH Equinoxe Narrow Modular Reverse Pilot-Tip Starter Reamer C is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Exactech, Inc.; Product: EXACTECH Equinoxe Narrow Modular Reverse Pilot-Tip Starter Reamer Catalog #: 321-45-01; Reason for recall: Instructions fo

Recalling firmExactech, Inc.
ProductEXACTECH Equinoxe Narrow Modular Reverse Pilot-Tip Starter Reamer Catalog #: 321-45-01
Reason for recallInstructions for use (IFU) is updated to clarify the proper technique and instrumentation to ream the Glenoid, i.e., to include a caution statement that the surgeon should ream on-axis and that reaming off-axis may result in instrument breakage.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionAL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico Australia, Austria, Brazil, Canada, China, France, Germany, Greece, India, Italy, Korea, Luxembourg, Singapore, South Africa, Spain, Sweden, Switzerland,
Recall initiated20170525
Recall numberZ-2667-2017
Categorydevice
StateFL
CountryUnited States