FHC, Inc. Recall: microTargeting" STar" Drive System (Manual), Product Number 70-ZD-MA

FHC, Inc. Recall: microTargeting" STar" Drive System (Manual), Product Number 70-ZD-MA is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: FHC, Inc.; Product: microTargeting" STar" Drive System (Manual), Product Number 70-ZD-MA; Reason for recall: The FHC Depth Stop Adapter may cause damag

Recalling firmFHC, Inc.
ProductmicroTargeting" STar" Drive System (Manual), Product Number 70-ZD-MA
Reason for recallThe FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionThe devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign c
Recall initiated20180510
Recall numberZ-2351-2018
Categorydevice
StateME
CountryUnited States