GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING Recall: Elscint SPX6; dual-head nuclear medi
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING Recall: Elscint SPX6; dual-head nuclear medi is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING; Product: Elscint SPX6; dual-head nuclear medicine gamma camera designed specifically for cardiac ima
| Recalling firm | GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING |
|---|---|
| Product | Elscint SPX6; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging |
| Reason for recall | GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | Worldwide |
| Recall initiated | 20250912 |
| Recall number | Z-0293-2026 |
| Category | device |
| Country | Israel |
$29/mo