GE Medical Systems, LLC Recall: Tec 850 SEV Anesthesia Vaporizer, Reference Number 1177-98

GE Medical Systems, LLC Recall: Tec 850 SEV Anesthesia Vaporizer, Reference Number 1177-98 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: GE Medical Systems, LLC; Product: Tec 850 SEV Anesthesia Vaporizer, Reference Number 1177-9850-SEV; Reason for recall: Due to a manufacturing issue

Recalling firmGE Medical Systems, LLC
ProductTec 850 SEV Anesthesia Vaporizer, Reference Number 1177-9850-SEV
Reason for recallDue to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionUS Nationwide. Global Distribution.
Recall initiated20241118
Recall numberZ-0818-2025
Categorydevice
StateWI
CountryUnited States