← RecallGuard
HomeRecalls

Getinge Group Logistics America, LLC Recall: Getinge CM320 Series Washer Disinfector - Pro

Getinge Group Logistics America, LLC Recall: Getinge CM320 Series Washer Disinfector - Pro is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Getinge Group Logistics America, LLC; Product: Getinge CM320 Series Washer Disinfector - Product Usage: A multi chamber washer disinfector intended

Recalling firmGetinge Group Logistics America, LLC
ProductGetinge CM320 Series Washer Disinfector - Product Usage: A multi chamber washer disinfector intended for washing, disinfecting and drying of surgical instruments, containers, operating theatre shoes and operating theatre accessories normally used in hospitals and central disinfection facilities. The disinfected items must be further processed through terminal sterilization before use in any surgical procedure.
Reason for recallThe potential that the plastic inlet hose connections may deform or break as a result of material fatigue and/or over tightening of the clamps. A broken hose connection may lead to a hazardous situation of water leakage or water spray. The water leakage or water spray may result in a burn to the user and/or an inconvenience due to a procedural delay if the Washer Disinfector is taken out of service.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide distribution - US Nationwide distribution including in the states of ND, PA, NC, VA, NE, MD, NY, MN and the countries of Australia, Austria, Belgium, Canada, Denmark, Finland, France, Germany, Hong Kong, Norway, Slovakia, Slovenia, South Korea, Sweden, Japan, China, Netherland, Italy, Kuwa
Recall initiated20201221
Recall numberZ-1124-2021
Categorydevice
StateNJ
CountryUnited States

🔍 Search all recalls →