Horiba Instruments, Inc dba Horiba Medical Recall: ABX Pentra Micro ALBUMIN 2 CP, Serum/Pl is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Horiba Instruments, Inc dba Horiba Medical; Product: ABX Pentra Micro ALBUMIN 2 CP, Serum/Plasma, REF 1300032563 The device is intended for use in

Recalling firmHoriba Instruments, Inc dba Horiba Medical
ProductABX Pentra Micro ALBUMIN 2 CP, Serum/Plasma, REF 1300032563 The device is intended for use in conjunction with certain materials to measure a variety of analytes.
Reason for recallWhen the ABX Pentra Micro ALBUMIN 2 CP (ALBT2) Reagent 2 is sampled and dispensed immediately after an ABX Pentra Creatinine 120 CP (CREA3) Reagent 2 dispense, the final result of the microalbumin may be elevated more than 10%. This increase may affect serum, plasma and urine sample types.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide Distribution: US (nationwide) and to countries of: Bulgaria, Burma, Cameroon, Djibouti, France, Gabon, Germany, Italy, Lebanon, Mali, Mauritania, Poland, Portugal, Romania, Saudi Arabia, South Africa and Thailand.
Recall initiated20180426
Recall numberZ-2286-2018
Categorydevice
StateCA
CountryUnited States