Horiba Instruments, Inc dba Horiba Medical Recall: Device Name: ABX PENTRA ML Model/Part #

Horiba Instruments, Inc dba Horiba Medical Recall: Device Name: ABX PENTRA ML Model/Part # is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Horiba Instruments, Inc dba Horiba Medical; Product: Device Name: ABX PENTRA ML Model/Part #(s): A11A01652 Lot/Serial #(s): N/A Software Version(s)

Recalling firmHoriba Instruments, Inc dba Horiba Medical
ProductDevice Name: ABX PENTRA ML Model/Part #(s): A11A01652 Lot/Serial #(s): N/A Software Version(s): All versions are subject to this correction. Product Usage: The ABX PENTRA ML is a software data management system intended for use on ABX Hematology Analyzers.
Reason for recallHORIBA Medical initiated this recall of all software versions of ABX PENTRA ML (Model A11A06152) due to a software bug with the potential of transmitting incorrect results to the Laboratory Information System.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionUSA Nationwide Distribution
Recall initiated20140214
Recall numberZ-1173-2014
Categorydevice
StateCA
CountryUnited States