Horiba Instruments Inc Recall: ABX PENTRA Glucose PAP CP ref. A11A01668 reagent is intende

Horiba Instruments Inc Recall: ABX PENTRA Glucose PAP CP ref. A11A01668 reagent is intende is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Horiba Instruments Inc; Product: ABX PENTRA Glucose PAP CP ref. A11A01668 reagent is intended for the quantitative in vitro diagnostic determinatio

Recalling firmHoriba Instruments Inc
ProductABX PENTRA Glucose PAP CP ref. A11A01668 reagent is intended for the quantitative in vitro diagnostic determination of glucose in human serum, plasma and urine using glucose oxidase method by colorimetry.
Reason for recallN-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionU.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.
Recall initiated20160111
Recall numberZ-0740-2016
Categorydevice
StateCA
CountryUnited States