Hospira Inc. Recall: LifeCare PCA Infusion System with Hospira MedNet Software, PCA Serial

Hospira Inc. Recall: LifeCare PCA Infusion System with Hospira MedNet Software, PCA Serial is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Hospira Inc.; Product: LifeCare PCA Infusion System with Hospira MedNet Software, PCA Serial List Number 20709 Allows for clinician administration

Recalling firmHospira Inc.
ProductLifeCare PCA Infusion System with Hospira MedNet Software, PCA Serial List Number 20709 Allows for clinician administration or self-administration of analgesic medications.
Reason for recallScrew rotation error occurring when the pump detects a mismatch between the number of steps given to the motor driving delivery and the number of steps the motor actually took. When this malfunction occurs, the pump will alarm and infusion stops.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionUnited States (Nationwide) Distribution including Puerto Rico and American Virgin Islands.
Recall initiated20130308
Recall numberZ-2873-2016
Categorydevice
StateIL
CountryUnited States