Hospira Inc. Recall: LifeCare PCA Infusion System with Hospira MedNet Software, PCA Serial
Hospira Inc. Recall: LifeCare PCA Infusion System with Hospira MedNet Software, PCA Serial is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Hospira Inc.; Product: LifeCare PCA Infusion System with Hospira MedNet Software, PCA Serial List Number 20709 Allows for clinician administration
| Recalling firm | Hospira Inc. |
|---|---|
| Product | LifeCare PCA Infusion System with Hospira MedNet Software, PCA Serial List Number 20709 Allows for clinician administration or self-administration of analgesic medications. |
| Reason for recall | Screw rotation error occurring when the pump detects a mismatch between the number of steps given to the motor driving delivery and the number of steps the motor actually took. When this malfunction occurs, the pump will alarm and infusion stops. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | United States (Nationwide) Distribution including Puerto Rico and American Virgin Islands. |
| Recall initiated | 20130308 |
| Recall number | Z-2873-2016 |
| Category | device |
| State | IL |
| Country | United States |