Icotec Ag Recall: VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702

Icotec Ag Recall: VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Icotec Ag; Product: VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702; Reason for recall: During internal testing (at manufactur

Recalling firmIcotec Ag
ProductVADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702
Reason for recallDuring internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionUS Nationwide distribution in the state of GA.
Recall initiated20221103
Recall numberZ-0728-2023
Categorydevice
CountrySwitzerland