Icotec Ag Recall: VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702
Icotec Ag Recall: VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Icotec Ag; Product: VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702; Reason for recall: During internal testing (at manufactur
| Recalling firm | Icotec Ag |
|---|---|
| Product | VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702 |
| Reason for recall | During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | US Nationwide distribution in the state of GA. |
| Recall initiated | 20221103 |
| Recall number | Z-0728-2023 |
| Category | device |
| Country | Switzerland |