ICU Medical, Inc. Recall: Product Description/Item: ChemoLock Bag Spike, 10 units/CL-10-10
ICU Medical, Inc. Recall: Product Description/Item: ChemoLock Bag Spike, 10 units/CL-10-10 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: ICU Medical, Inc.; Product: Product Description/Item: ChemoLock Bag Spike, 10 units/CL-10-10; ChemoLock BAG SPIKE w/ ADDITIVE PORT, DRY SPIKE/CL-12
| Recalling firm | ICU Medical, Inc. |
|---|---|
| Product | Product Description/Item: ChemoLock Bag Spike, 10 units/CL-10-10; ChemoLock BAG SPIKE w/ ADDITIVE PORT, DRY SPIKE/CL-12; APPX 1.5ml,ChemoLock BAG SPIKE WITH ADDITIVE PORT/CL-13; ChemoLock Port/CL2100; ChemoLock Port, 5 Units/CL2100-5; ChemoLock Port/CL2100T; Clave Bag Spike w ChemoLock Port/CL2150; 30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger/CL3011; 9 IN(23cm)APPX 2.7ml,EXT SET,ChemoLock PORT,2 CLAMPS,2 GRADUATED CONN/CL3364; Syringe Transfer Set w Clave, ChemoLock Port/CL-34; 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w Integrated C |
| Reason for recall | Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | Worldwide distribution - US Nationwide distribution including in the states of LA, FL, TX, NY, NE, WI, PA, IA, ND, CA, OH, WA, GA, AL, CO, AZ, NC, TN, MI, VA, MA, DE, OR, KY, CT, OK, NJ, MN, IL, VT, AR, RI, ID, IN, UT, NV, MT, MD, KS, SC, NM, ME, MO, SD and the countries of : Saudi Arabia, Australia |
| Recall initiated | 20220627 |
| Recall number | Z-1681-2022 |
| Category | device |
| State | CA |
| Country | United States |