Infusion Pump Repair Recall: Alaris Pump Module, Model 8100, serviced/repaired by Infusion

Infusion Pump Repair Recall: Alaris Pump Module, Model 8100, serviced/repaired by Infusion is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Infusion Pump Repair; Product: Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part; Reason

Recalling firmInfusion Pump Repair
ProductAlaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part
Reason for recallBezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
ClassificationClass I (most serious — reasonable probability of serious health consequences or death)
StatusOngoing
DistributionCalifornia
Recall initiated20210326
Recall numberZ-1544-2021
Categorydevice
StateCA
CountryUnited States