Inpeco S.A. Recall: FlexLab, Accelerator a3600 and Aptio Automation are modular systems de
Inpeco S.A. Recall: FlexLab, Accelerator a3600 and Aptio Automation are modular systems de is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Inpeco S.A.; Product: FlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre-Analytical and Post-Analytical p
| Recalling firm | Inpeco S.A. |
|---|---|
| Product | FlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation. |
| Reason for recall | Inpeco received several notifications from the field about burnt schuko sockets. The customers reported that some smoke was seen coming from the damaged sockets. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Worldwide Distribution - US including Texas and New York. Foreign Consignees: Germany, Italy, France, Austria and Spain. |
| Recall initiated | 20161004 |
| Recall number | Z-0587-2017 |
| Category | device |
| State | N/A |
| Country | Switzerland |