Inpeco S.A. Recall: FlexLab, Accelerator a3600 and Aptio Automation are modular systems de

Inpeco S.A. Recall: FlexLab, Accelerator a3600 and Aptio Automation are modular systems de is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Inpeco S.A.; Product: FlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre-Analytical and Post-Analytical p

Recalling firmInpeco S.A.
ProductFlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.
Reason for recallInpeco received several notifications from the field about burnt schuko sockets. The customers reported that some smoke was seen coming from the damaged sockets.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide Distribution - US including Texas and New York. Foreign Consignees: Germany, Italy, France, Austria and Spain.
Recall initiated20161004
Recall numberZ-0587-2017
Categorydevice
StateN/A
CountrySwitzerland

🔍 Search all recalls →