Intuitive Surgical, Inc. Recall: da Vinci X (IS4200) and Xi (IS4000) systems, which consis

Intuitive Surgical, Inc. Recall: da Vinci X (IS4200) and Xi (IS4000) systems, which consis is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Intuitive Surgical, Inc.; Product: da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon Console (SSC), Patient C

Recalling firmIntuitive Surgical, Inc.
Productda Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon Console (SSC), Patient Cart (PSCART), Vision System (VSS)
Reason for recallInadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3) Yellow pedal associated with force bipolar is pressed, released with head in surgeon console 4) Blue pedal associated with bipolar is pressed, resulting in force bipolar energy delivery
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionUS: MS, PA, WI, KY, TN, MI, FL, GA, NC, NV, LA, OH, IL, MO, IN, CA, AK, MD, TX, IA, MA, NE, AZ, WA, CO, AL, SC
Recall initiated20220920
Recall numberZ-0133-2023
Categorydevice
StateCA