← RecallGuard
HomeRecalls

Landauer Recall: Calibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 18120-000; b) 1812

Landauer Recall: Calibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 18120-000; b) 1812 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Landauer; Product: Calibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 18120-000; b) 18120-KIT; c) 18120-SET; radiation monitoring dosimeter used with

Recalling firmLandauer
ProductCalibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 18120-000; b) 18120-KIT; c) 18120-SET; radiation monitoring dosimeter used with the microSTAR readers
Reason for recallLANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusCompleted
DistributionWorldwide
Recall initiated20230712
Recall numberZ-2564-2023
Categorydevice
StateIL
CountryUnited States

🔍 Search all recalls →