← RecallGuard
HomeRecalls

Landauer Recall: nanoDot D2DNS, Model Numbers: a) 03057-1MO; b) 03057-3MO; c) 03057-OTO

Landauer Recall: nanoDot D2DNS, Model Numbers: a) 03057-1MO; b) 03057-3MO; c) 03057-OTO is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Landauer; Product: nanoDot D2DNS, Model Numbers: a) 03057-1MO; b) 03057-3MO; c) 03057-OTO; d) 03057-SMO; radiation monitoring dosimeter used with the

Recalling firmLandauer
ProductnanoDot D2DNS, Model Numbers: a) 03057-1MO; b) 03057-3MO; c) 03057-OTO; d) 03057-SMO; radiation monitoring dosimeter used with the microSTAR readers
Reason for recallLANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusCompleted
DistributionWorldwide
Recall initiated20230712
Recall numberZ-2544-2023
Categorydevice
StateIL
CountryUnited States

🔍 Search all recalls →