← RecallGuard
HomeRecalls

Life Technologies Corporation Recall: Torrent Suite Dx Software versions 5.14 and earlier

Life Technologies Corporation Recall: Torrent Suite Dx Software versions 5.14 and earlier is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Life Technologies Corporation; Product: Torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems-IVD intended

Recalling firmLife Technologies Corporation
ProductTorrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems-IVD intended for targeted sequencing of human genomic DNA (gDNA) from peripheral whole-blood samples and DNA and RNA extracted from formalin-fixed, paraffin-embedded (FFPE) samples. The Ion PGM Dx Instrument System is not intended for whole genome or de novo sequencing Product Name Installer /SKU /Updater SKU: (1) Torrent Suite Dx Software 5.14 -A52422/ A52417 (2) Torrent Suite Dx Software 5.12.5- A46166/ A46167 (3)Torrent Suite Dx Software 5.8- A36601/ A36602 (4)Torrent Suite Dx Software 5.6.4- A331
Reason for recallTorrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems exploitation of the vulnerability by a threat actor may allow them to alter settings, configurations, software, or data on the instrument
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionUS Nationwide distribution in the states of CA, IN, NC OR, TN.
Recall initiated20240207
Recall numberZ-1426-2024
Categorydevice
StateMD
CountryUnited States

🔍 Search all recalls →