Lin-Zhi International Inc Recall: Cannabinoids DAU Calibrator Control Level 1 (62.5 ng/mL)
Lin-Zhi International Inc Recall: Cannabinoids DAU Calibrator Control Level 1 (62.5 ng/mL) is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Lin-Zhi International Inc; Product: Cannabinoids DAU Calibrator Control Level 1 (62.5 ng/mL) This solution is prepared from non-sterile urine and s
| Recalling firm | Lin-Zhi International Inc |
|---|---|
| Product | Cannabinoids DAU Calibrator Control Level 1 (62.5 ng/mL) This solution is prepared from non-sterile urine and should be handled as potentially infectious. IVD; The Cannabinoids (cTHC also known as THC version c) DAU negative and positive controls are human urine-based liquids and ready to use. The constituent is a processed drug-free human urine matrix, containing buffers, stabilizers, and less than 0.09 % of sodium azide. The controls are prepared by spiking known concentrations of 11-nor-9-THC-9-COOH into the drug-free matrix. cTHC is referenced as an abbreviation for 11-nor-9-THC-9-COOH a |
| Reason for recall | Product catalog #s 0007c Cannabinoid 37.5 ng/mL Control Level 1 and 0008c Cannabinoid 62.5 Control Level 2 of the lots mentioned may give decreased concentration readings towards the end of their shelf life due to use of dated but unexpired secondary stock used to create the products. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | US Distribution including the states of SC, NC, LA, FL, TX,, AL and GA and Internationally to Germany. |
| Recall initiated | 20140206 |
| Recall number | Z-1179-2014 |
| Category | device |
| State | CA |
| Country | United States |