Maquet Cardiovascular, LLC Recall: Maquet Getinge- (1)Material: 701065765R01 BO-TOP 36504

Maquet Cardiovascular, LLC Recall: Maquet Getinge- (1)Material: 701065765R01 BO-TOP 36504 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Maquet Cardiovascular, LLC; Product: Maquet Getinge- (1)Material: 701065765R01 BO-TOP 36504 Adult ECC Pack (2)Material: 701065765R02 BO-TOP 36504 Ad

Recalling firmMaquet Cardiovascular, LLC
ProductMaquet Getinge- (1)Material: 701065765R01 BO-TOP 36504 Adult ECC Pack (2)Material: 701065765R02 BO-TOP 36504 Adult ECC Pack
Reason for recallCustom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionNationwide
Recall initiated20190418
Recall numberZ-1663-2019
Categorydevice
StateNJ
CountryUnited States