Maquet Cardiovascular Us Sales, Llc Recall: Magnus Hybrid OR Table Column (1180.01B2) used
Maquet Cardiovascular Us Sales, Llc Recall: Magnus Hybrid OR Table Column (1180.01B2) used is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Maquet Cardiovascular Us Sales, Llc; Product: Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography Sy
| Recalling firm | Maquet Cardiovascular Us Sales, Llc |
|---|---|
| Product | Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co. |
| Reason for recall | There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Worldwide Distribution - US Nationwide Distribution and to the countries of : France, Australia, Belgium, Brazil, Switzerland, Chile, China, Germany, Spain, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Japan, Kuwait, Kazakhstan, Mexico, Netherlands, Norway, Russia, Sweden, Turkey, Taiwa |
| Recall initiated | 20150610 |
| Recall number | Z-2097-2015 |
| Category | device |
| State | NJ |
| Country | United States |