Maquet Cardiovascular Us Sales, Llc Recall: Magnus Hybrid OR Table Column (1180.01B2) used

Maquet Cardiovascular Us Sales, Llc Recall: Magnus Hybrid OR Table Column (1180.01B2) used is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Maquet Cardiovascular Us Sales, Llc; Product: Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography Sy

Recalling firmMaquet Cardiovascular Us Sales, Llc
ProductMagnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co.
Reason for recallThere is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide Distribution - US Nationwide Distribution and to the countries of : France, Australia, Belgium, Brazil, Switzerland, Chile, China, Germany, Spain, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Japan, Kuwait, Kazakhstan, Mexico, Netherlands, Norway, Russia, Sweden, Turkey, Taiwa
Recall initiated20150610
Recall numberZ-2097-2015
Categorydevice
StateNJ
CountryUnited States