MAQUET GMBH Recall: Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3)
MAQUET GMBH Recall: Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: MAQUET GMBH; Product: Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All
| Recalling firm | MAQUET GMBH |
|---|---|
| Product | Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions. |
| Reason for recall | Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | Worldwide - US Nationwide distribution in the states of KY, NY, TX and the countries of Algeria, Australia, Austria, Bulgaria, Canada, China, Colombia, Denmark, Egypt, Estonia, Finland, France, Germany, Ireland, Israel, Italy, Japan, Jordan, Korea, Republic of, Kuwait, Lebanon, Morocco, Netherlands, |
| Recall initiated | 20250714 |
| Recall number | Z-2420-2025 |
| Category | device |
| Country | Germany |