Medical Components, Inc dba MedComp Recall: 12F Tesio CATHETER SET (CUFF 32CM FROM TIP), R
Medical Components, Inc dba MedComp Recall: 12F Tesio CATHETER SET (CUFF 32CM FROM TIP), R is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Medical Components, Inc dba MedComp; Product: 12F Tesio CATHETER SET (CUFF 32CM FROM TIP), REF MCTC1247SD, UDI 884908027266 Product Usage: The Medc
| Recalling firm | Medical Components, Inc dba MedComp |
|---|---|
| Product | 12F Tesio CATHETER SET (CUFF 32CM FROM TIP), REF MCTC1247SD, UDI 884908027266 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing. |
| Reason for recall | There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland |
| Recall initiated | 20180223 |
| Recall number | Z-1124-2018 |
| Category | device |
| State | PA |
| Country | United States |