Medicrea International Recall: UNiD" Adaptive Spine Intelligence UNiD Spine Analyzer Produ is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Medicrea International; Product: UNiD" Adaptive Spine Intelligence UNiD Spine Analyzer Product Number SW3002 UDI-DI code 03613720286929 Two cloud-b

Recalling firmMedicrea International
ProductUNiD" Adaptive Spine Intelligence UNiD Spine Analyzer Product Number SW3002 UDI-DI code 03613720286929 Two cloud-based software applications within UNiD Adaptive Spine Intelligence (ASI): the UNiD HUB and the UNiD Spine Analyzer. The UNiD HUB cloud-based software is a healthcare application to receive, transfer, display, store data needed for planning a spine surgery or for post-operative follow-up (patient information, X-ray image and recommendations for planning). The UNiD Spine Analyzer cloud based software is a healthcare application intended for assisting healthcare professionals in viewi
Reason for recallDue to software anomalies that may impact on rod planning specifically, certain optional surgical parameters may have had errors that resulted i incorrect calculations displayed on system. Impacted parameters includes the following: Roussouly Classification, Real Lumbar Lordosis and Real Thoracic Kyphosis, Barrey Ratio, and Lenke Classification.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionU.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. O.U.S.: Argentina Australia Belgium Brazil Bulgaria Chile Czech Republic Denmark Finland F
Recall initiated20241203
Recall numberZ-1118-2025
Categorydevice
CountryFrance