← RecallGuard
HomeRecalls

MEDLINE INDUSTRIES, LP - Northfield Recall: Centurion Kits, trays, and packs labeled as fo

MEDLINE INDUSTRIES, LP - Northfield Recall: Centurion Kits, trays, and packs labeled as fo is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: MEDLINE INDUSTRIES, LP - Northfield; Product: Centurion Kits, trays, and packs labeled as follows: a) IV SECUREMENT KIT, REF IVS1860B; b) NEONATAL

Recalling firmMEDLINE INDUSTRIES, LP - Northfield
ProductCenturion Kits, trays, and packs labeled as follows: a) IV SECUREMENT KIT, REF IVS1860B; b) NEONATAL PICC LINE INSERTION TRAY, REF CVI2025; c) NEONATAL PICC LINE INSERTION TRAY, REF CVI4570; d) SECONDARY IV START KIT, REF IV8695A
Reason for recallMedline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
ClassificationClass I (most serious — reasonable probability of serious health consequences or death)
StatusOngoing
DistributionDistribution US nationwide including Puerto Rico, Canada, UAE, and Panama
Recall initiated20231115
Recall numberZ-1134-2024
Categorydevice
StateIL
CountryUnited States

🔍 Search all recalls →