Medline Industries, LP Recall: Medline Convenience Kits: 1) LITHOTOMY PACK, Model Number:
Medline Industries, LP Recall: Medline Convenience Kits: 1) LITHOTOMY PACK, Model Number: is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Medline Industries, LP; Product: Medline Convenience Kits: 1) LITHOTOMY PACK, Model Number: DYNJ83185; 2) LITHOTOMY PACK, Model Number: DYNJ83185A;
| Recalling firm | Medline Industries, LP |
|---|---|
| Product | Medline Convenience Kits: 1) LITHOTOMY PACK, Model Number: DYNJ83185; 2) LITHOTOMY PACK, Model Number: DYNJ83185A; 3) LAP BASIN, Model Number: DYNJVB91109D |
| Reason for recall | Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | Worldwide - US Nationwide distribution. |
| Recall initiated | 20260107 |
| Recall number | Z-2100-2026 |
| Category | device |
| State | IL |
| Country | United States |